GE-MON

Central monitoring and alarm management for critical equipment and environments: ULT freezers, cryogenic tanks, incubators, cleanrooms and infrastructure, with a staged alarm chain and audit-ready reporting.

Available

A minus eighty freezer usually fails at night or over a weekend. Its own audible alarm sounds in an empty corridor, and the first person in on Monday finds years of samples thawed. Central monitoring exists for that scenario, and the answer is not simply "add a sensor": the right threshold, the right delay, an escalation path that reaches a person, and a record that can later serve as evidence are all required together.

What is monitored

AreaParameters
Freezers (+4 to −86 °C)Temperature, door contact, device alarm contact, loss of device communication
Cryogenic tanksInternal temperature, liquid nitrogen level, fill status, controller alarm
IncubatorsTemperature, CO₂, O₂, relative humidity, door
Cleanroom and laboratoryTemperature, humidity, differential pressure, particle count, cabinet airflow
Cryogenic storage roomOxygen concentration, visual and audible warning, HVAC relay control
InfrastructureUPS and generator status, water leak, medical gas and LN2 supply line
BIOSECURE environmental monitoring dashboard, list view: six devices in the GMP Lab group — LN2 tank at −162.31 °C, ultra-low freezer at −70.54 °C, incubator (84.67% humidity and 36.15 °C), a second incubator at 33.17 °C, a LAF cabinet at 25.42 °C and a Telstar unit at 25.32 °C; humidity and temperature trends for the selected device on the right
One screen covers freezers, LN2 tanks, incubators and clean-room cabinets; each card opens its own trend.

Staged alarms

Threshold, delay and escalation path are configured independently for each device. A typical chain waits for a defined period, then sends email, then SMS, then places a voice call to a responsible list. Who acknowledged the alarm, when and with what reason is part of the record — alarm delays should be justified and documented, not left implicit.

BIOSECURE alarm history: timeline of a high-temperature alarm on a −86 °C freezer — alarm raised at 09:42, e-mail sent at 09:46, SMS at 09:50, phone call at 09:53, alarm cleared at 10:37, 54 minutes in total. Fault and power-interruption records appear below
Alarm history shows which notification went out when, and how long the alarm stayed open.

How data is collected

  • GE-202 for temperature, humidity and contact inputs over MQTT
  • GE-100 for oxygen, carbon dioxide and differential pressure over Modbus
  • GE-EDGE for existing controllers from PHCbi, Haier, Worthington, Chart, Nüve and others
  • Existing monitoring hardware from third parties can be adopted rather than replaced

Records and reporting

Measurements are stored in a time series database with device, sensor and timestamp. Dashboards, historical charts with threshold bands, excursion and recovery curves, and PDF or Excel reports for a chosen period are available for audit. Access is role-based and changes are written to an audit trail.

Calibration certificates and maintenance records are attached to the device card, so an inspector can see the measurement and its traceability in the same place.

Talk to our team

Send your equipment list and we will propose an architecture and sensor plan.